Specifications:
System Function:
Genearl : Benchtop,sutomatic,Distrete,Random Access,STAT Sample priority
Throughput: Up to 120 tests per hour
Princeple: Colorimetry and turbidimetry
Methodology:End-point,fixed-tim,kinetic,etc
Programming:Open or closed ystem on demand
Reagent/Sample Handling :
Reagent/Sample Tray : 54 Reagent Positions 27 Sample Positions
Reagent Volume : 10-300ul,Step by 1 ul
Sample Volume : 1-70ul,step by 0.1ul
Reagent/Sample Probe : Liquid Level Detection Vertical&horizontal collision detection and reagents inventory monitoring
Probe Cleaning : Auto interior and exterior wash
Reaction system:
Reaction tray:81 reusable cuvettes,auto cuvette wash
Reaction volume:150-750ul
Reaction temperature:37±0.1ºC
Heating Method:Solid Heating
Optical System:
Light Source:Halogen Tungsten Lamp
Photometer:Maintenance-Free Photometer,rear spectrophotometry by filters
Wavelengh:8 wavelengths:340nm,405nm,450nm,510nm,546nm,578nm,630nm,670nm
Absorbance;0-4.5Abs
Calibration and control:
Calibration:K-factor,Linear,Spline,Logit-log 4p,Logit-Log 5p,Exponential,Polynomial
Control: Westgard multi-rule,Cumulative sum check,Twin plot,L-J Chart
Operation System:
Computer: Built-in operation system
Screen:10.4 inch color touch screen
Data Storage:100,000 results
Printer:Built-in and external printer
Others:Soft keyboard
compatible with keyboard and mouse
Interface:Ethernet,USB,RS232,VGA,HDMI,Audio,etc
Working ConditionL
Poer Supply:AC100~240V,50/60Hz,≤150VA
Temperature:10-30ºC
Humidity:30-85%RH
Dimension:605mm*350mm*350mm(L*W*H)
Weight:25kgs
Water Consumption: ≤1L/H